Fiber Spectrometers GLA600-UVN

VIS fiber spectrometer

ModelGLA600-UVN
BrandGlit
PlaceShenzhen

  • Czerny-Turner optical structure with a grating, CCD detector and a SMA905 fiber connector.

    Connects to a computer through USB, controlled and powered by the computer.

    Unique software features, automatic configuration of spectrometer, automatic reading of calibration coefficients, spectral acquisition and measurement, logging and color bar display.

    Spectral measurements including radiation, transmittance, absorbance and reflectivity.

    Two output data formats, conventional Excel two-column format and square format, the latter can be used to check spectral data easily which is a unique design developed by us.

    The selected spectral data on the spectral curve are marked using different colors, including the cursors and the area boundary lines, allowing users to easily find data in a big amount of the spectral data.

    Cursors are used to find the wavelengths. The spectral width is found by moving the two boundary lines.

    Software logging allows the user to record experiment’s content such as the material analyzed, operator and time etc.

    Color bars help users understand the corresponding wavelengths of the spectrum.

    GLA600-UVN fiber spectrometer includes UV bands spectrometer, VIS band spectrometer and NIR band spectrometer.  


  • Excellent Consistency of mass production for industrial applications  

    Excellent consistency of thermal stability and vibration resistance

    Compact size and sleek appearance.


  • Items Specifications Remarks
    Fiber Connector SMA905
    Slit 10, 25, 50, 100µm
    Wavelength Range 350-1000nm Depends on the grating used
    Spectral Resolution FWHM 0.76nm@576nm@25µm slit Depends on wavelengths and slits used
    Order Sorting Filter To eliminate higher order spectrum
    Stray Light 0.05%@546nm Depends on wavelengths
    Light Detector Toshiba TCD1304DG
    Pixel Number 3648
    Pixel Size 8μmx200μm
    Detector Collecting Lens Yes
    SNR 800:1 Room temperature
    ADC Resolution 16 bits
    Integration Time 8 ms-15min Depends on the detector used
    Data Output Count vs. wavelength or pixel
    Power Supply DC5V/250mA
    Communication USB
    Operation Software SPEC-GLA600
    Application Programming Interface glaDevSys For users to develop software
    Dimensions 100mmx65mmx36mm
    Weight 285g
    Mounting 4xM3 screw holes on the bottom
    Operating Temperature -10-50
    Storage Temperature -20-70

  • Environmental inspection: inspection of water quality and air pollution etc.

    Chemical industry  

    Optical coating inspection

    Glass and transparent materials inspection

    Measuring light sources

    R&D in science and technology

    Customization